Detailed instructions for use are in the User's Guide.
[. . . ] Digital Multimeters
Tektronix DMM4020 Data Sheet
Available Functions and Features
Volts, Ohms, and Amps Measurements Diode and Continuity Testing Frequency Measurements 2×4 Ohms 4-wire Measurement Technique Dedicated DC Leakage Current Measurements Six Dedicated Buttons for Fast Access to Instrument Setups Limit Compare Mode for Pass/Fail Testing
Connectivity
Front 2×4 Measurement Inputs
DMM4020
RS-232 on Rear Panel for Quick PC Connectivity Includes USB to RS-232 Interface Adapter Cable Includes National Instrument's LabVIEW SignalExpressTM TE Limited Edition for Connecting Your Bench
Features & Benefits
Key Performance Specifications
5. 5 Digit Resolution Basic VDC Accuracy of up to 0. 015% (1 yr. ) 200 mV to 1000 V Voltage Range, with up to 1 V Resolution 200 A to 10 A Current Range, with up to 1 nA Resolution 200 to 100 M Range, with up to 1 m Resolution CAT I 1000 V, CAT II 600 V
3-year Warranty
Data Sheet
Measurements with the Push of a Button
As the circuits in embedded system designs become more sophisticated, you must measure a multitude of different parameters to validate your design. The Tektronix DMM4020 5. 5 digit bench multimeter offers a broad range of functions in one easy-to-use instrument. Typical multimeter measurements volts, ohms, and amps are made with a basic VDC accuracy of up to 0. 015%, ensuring you have the performance you need for your design. [. . . ] Special test lead accessories are available to enable you to establish the connection. You get excellent resolution and accuracy plus the convenience and ease of using a single pair of leads. Easy Connectivity to Your PC The RS-232 port on the back panel can be used to connect to your PC. A USB to RS-232 interface adapter cable is included standard with the DMM4020 for connecting to your PC's USB port.
Designed to Make Your Work Easier
The DMM4020 multimeter is designed with the ease of use and familiar operation you expect from Tektronix.
Intuitive Operation
Dedicated front-panel buttons provide fast access to frequently used functions and parameters, reducing setup time. You no longer need to search through software menus to find the function you need.
2
www. tektronix. com/dmm
Digital Multimeters -- Tektronix DMM4020
SignalExpress acquiring data from Tektronix DMM4020 and DPO3052.
SignalExpress used to export DMM4020 data into Excel.
Connect Your Bench for Intelligent Debug
Easily capture, save, and analyze measurement results from your multimeter with the special Tektronix Edition of National Instrument's LabVIEW SignalExpressTM software. Every DMM4020 multimeter ships with a free copy of the Limited Edition version of SignalExpress for basic instrument control, data logging, and analysis. The optional Professional Edition offers over 200 built-in functions that provide additional signal processing, advanced analysis, sweeping, limit testing, and user-defined step capabilities. SignalExpress supports the range of Tektronix bench instruments*1, enabling you to connect your entire test bench. You can then access the
feature-rich tools packed into each instrument from one intuitive software interface. This allows you to automate complex measurements requiring multiple instruments, log data for an extended period of time, time correlate data from multiple instruments, and easily capture and analyze your results, all from your PC. Only Tektronix offers a connected test bench of intelligent instruments to simplify and speed debug of your complex design.
Performance You Can Count On
In addition to industry-leading service and support, every DMM4020 multimeter comes backed with a three-year standard warranty.
*1 NI LabVIEW SignalExpress supports the following Tektronix instruments: MSO/DPO4000/3000/2000 Series oscilloscopes, TDS3000C/2000B/1000B Series oscilloscopes, AFG3000 Series arbitrary/function generators, DMM4050/4040/4020 Series digital multimeters.
www. tektronix. com/dmm
3
Data Sheet
Characteristics
General Specifications
Voltage
Characteristic Description
Triggering
Characteristic Description
100V Setting 120V Setting 220V Setting 240V Setting Frequency Power Consumption
90 V to 110 V 108 V to 132 V 198 V to 242 V 216 V to 264 V 47 Hz to 440 Hz 15 VA peak (10 W average)
Trigger Delay External Trigger Delay External Trigger Jitter Trigger Input Trigger Output
400 ms <2 ms <1 ms TTL Levels 5 V max
Math Functions
Min/Max, relative, hold, compare, and dB functions
in.
Dimensions
Dimension mm
Height Width Depth
Weight
88 217 297
kg
3. 46 8. 56 11. 7
lb.
Electrical
Characteristic Description
Input Protection Overrange
Net
2. 1
4. 6
1000 V all ranges 10% on the largest ranges of all functions except continuity and diode test
Display
Vacuum Fluorescent Display, segment
Remote Interfaces
RS-232C (RS-232 to USB Adapter cable included)
Environment
Characteristic Description
Warranty
Three years
Temperature Operating 0 °C to 50 °C Storage 40 °C to 70 °C Warm Up ½ hour to full uncertainty specifications Relative Humidity (noncondensing) Operating Uncontrolled (<10 °C) <90% (10 °C to 28 °C) <75% (28 °C to 40 °C) <45% (40 °C to 50 °C) Storage <95% (40 °C to 70 °C) Altitude 2, 000 meters Operating 12, 000 meters Storage Vibration Complies with MIL-PRF-28800F Class 3 Safety Complies with IEC 61010-1:2001, ANSI/ISA 61010-1 (S82. 02. 01):2004, UL 61010-1:2004, CAN/CSA C22. 2 No. 61010. 1:2004, CAT I 1000V / CAT II 600 V EMC Designed to comply with IEC 61326-1:1997+A1:1998+A2:2000
4
www. tektronix. com/dmm
Digital Multimeters -- Tektronix DMM4020
Electrical Specifications DC Voltage Specifications
Specifications are valid for 5½ digit mode and after at least a half-hour warm up.
Characteristic Description
Maximum Input Common Mode Rejection Normal Mode Rejection A/D Nonlinearity Input Bias Current Settling Considerations
1000 V on any range 120 dB at 50 or 60 Hz 0. 1% (1 k unbalance) 80 dB at Slow Rate 15 ppm of range <30 pA at 25 °C Measurement settling times are affected by source impedance, cable dielectric characteristics, and input signal changes
Input Characteristics
Range Full Scale (5½ Digits) Resolution Slow Medium Fast Input Impedance
200 mV 2V 20 V 200 V 1000 V
199. 999 mV 1. 99999 V 19. 9999 V 199. 999 V 1000. 00 V
1 V 10 V 100 V 1 mV 10 mV
10 V 100 V 1000 V 10 mV 100 mV
10 V 100 V 1000 V 10 mV 100 mV
>10 G*2 >10 G*2 10 M ±1% 10 M ±1% 10 M ±1%
*2 At some dual-display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 M.
Accuracy
Range 90 days 23 °C ±5 °C Uncertainty*3 1 year 23 °C ±5 °C Temperature Coefficient/°C Outside 18 - 28 °C
200 mV 2V 20 V 200 V 1000 V
*3 Uncertainty given as ±(% of reading + % of range).
0. 01 0. 01 0. 01 0. 01 0. 01
+ + + + +
0. 003 0. 002 0. 003 0. 002 0. 002
0. 015 0. 015 0. 015 0. 015 0. 015
+ + + + +
0. 004 0. 003 0. 004 0. 003 0. 003
0. 0015 + 0. 0005 0. 001 + 0. 0005 0. 0020 + 0. 0005 0. 0015 + 0. 0005 0. 0015 + 0. 0005
www. tektronix. com/dmm
5
Data Sheet
AC Voltage Specifications
AC Voltage specifications are for AC sinewave signals >5% of range. For inputs from 1% to 5% of range and <50 kHz, add an additional error of 0. 1% of range, and for 50 kHz to 100 kHz, add 0. 13% of range.
Characteristic Description
Maximum Input Measurement Method AC Filter Bandwidth Common Mode Rejection Maximum Crest Factor Additional Crest Factor Errors (<100 Hz)
750 VRMS or 1000 V peak or 8 × 107 Volts-Hertz product AC-coupled true RMS. Measures the AC component of input with up to 1000 VDC bias on any range 20 Hz - 100 kHz 60 dB at 50 Hz or 60 Hz (1 k unbalance) 3:1 at Full Scale Crest Factor 1-2, 0. 05% of full scale Crest Factor 2-3, 0. 2% of full scale Only applies for non-sinusoid signals
Input Characteristics
Range Full Scale (5½ Digits) Resolution Slow Medium Fast Input Impedance
200 mV 2V 20 V 200 V 750 V
199. 999 mV 1. 99999 V 19. 9999 V 199. 999 V 750. 00 V
1 V 10 V 100 V 1 mV 10 mV
10 V 100 V 1000 V 10 mV 100 mV
10 V 100 V 1000 V 10 mV 100 mV
1 M ±2% shunted by <100 pf
Accuracy
Range Frequency 90 days 23 °C ±5 °C Uncertainty*3 1 year 23 °C ±5 °C Temperature Coefficient/°C Outside 18 - 28 °C
200 mV
2V
20 V
200 V
750 V
20 Hz - 45 Hz 45 Hz - 20 kHz 20 kHz - 50 kHz 50 kHz - 100 kHz 20 Hz - 45 Hz 45 Hz - 20 kHz 20 kHz - 50 kHz 50 kHz - 100 kHz 20 Hz - 45 Hz 45 Hz - 20 kHz 20 kHz - 50 kHz 50 kHz - 100 kHz 20 Hz - 45 Hz 45 Hz - 20 kHz 20 kHz - 50 kHz 50 kHz - 100 kHz 20 Hz - 45 Hz 45 Hz - 20 kHz 20 kHz - 50 kHz 50 kHz - 100 kHz
0. 8 + 0. 05 0. 15 + 0. 05 0. 3 + 0. 05 0. 8 + 0. 05 0. 8 + 0. 05 0. 15 + 0. 05 0. 3 + 0. 05 0. 8 + 0. 05 0. 8 + 0. 05 0. 15 + 0. 05 0. 3 + 0. 05 0. 8 + 0. 05 0. 8 + 0. 05 0. 15 + 0. 05 0. 3 + 0. 05 0. 8 + 0. 05 0. 8 + 0. 05 0. 15 + 0. 05 0. 3 + 0. 05 0. 8 + 0. 05
0. 9 + 0. 05 0. 2 + 0. 05 0. 35 + 0. 05 0. 9 + 0. 05 0. 9 + 0. 05 0. 2 + 0. 05 0. 35 + 0. 05 0. 9 + 0. 05 0. 9 + 0. 05 0. 2 + 0. 05 0. 35 + 0. 05 0. 9 + 0. 05 0. 9 + 0. 05 0. 2 + 0. 05 0. 35 + 0. 05 0. 9 + 0. 05 0. 9 + 0. 05 0. 2 + 0. 05 0. 35 + 0. 05 0. 9 + 0. 05
0. 01 + 0. 005 0. 01 + 0. 005 0. 01 + 0. 005 0. 05 + 0. 01 0. 01 + 0. 005 0. 01 + 0. 005 0. 01 + 0. 005 0. 05 + 0. 01 0. 01 + 0. 005 0. 01 + 0. 005 0. 01 + 0. 005 0. 05 + 0. 01 0. 01 + 0. 005 0. 01 + 0. 005 0. 01 + 0. 005 0. 05 + 0. 01 0. 01 + 0. 005 0. 01 + 0. 005 0. 01 + 0. 005 0. 05 + 0. 01
*3 Uncertainty given as ±(% of reading + % of range).
6
www. tektronix. com/dmm
Digital Multimeters -- Tektronix DMM4020
Resistance
Specifications are for 4-wire resistance function, or 2-wire resistance with REL. [. . . ] Tektronix maintains a comprehensive, constantly expanding collection of application notes, technical briefs and other resources to help engineers working on the cutting edge of technology. Information in this publication supersedes that in all previously published material. All other trade names referenced are the service marks, trademarks, or registered trademarks of their respective companies. [. . . ]